Product Name: FCDS-02R Air Disinfector
Application: it is suitable for air disinfection in outpatient rooms, offices, examination rooms, treatment rooms, infectious diseases clinics and other places.
Technical parameter requirements:
1. Plasma + active mist ion is used for disinfection and sterilization, which is broad-spectrum and thorough; It contains HEPA high-efficiency filter, which can effectively remove volatile gases and various odors in the air and filter large dust particles such as hair and dust;
2. External dimension: 300 mm ×φ 150 mm, non removable, desktop equipment;
3. Human machine coexistence, continuous dynamic disinfection can be carried out in the presence of people, without any harm to people and objects;
4. Overall weight ≤ 1.3KG, rated circulating air volume ≥ 32m ³/ h. Applicable to 20m ³ Places with volume and below;
5. Rated power: 15W; Power supply 220V 50Hz;
★ 6. Equipped with active mist ion generation movement, the hydroxyl radical release efficiency of the active mist ion generation movement shall not be less than 6.9 × 1011 / s, the life of hydroxyl radicals in the released active mist ions is not less than 8 minutes;
★ 7. The safety of hydroxyl radicals generated by the active mist ion generation module is verified by acute inhalation toxicity test;
8. The equipment works continuously for 1 hour, and the ozone residue is less than 0.09mg/m ³。
★ 9. The equipment is equipped with 4G / 5G communication chips, which can be connected to the Internet of things equipment management platform. After logging in to the management platform, you can check the equipment operation status and the service life of consumables online, and can remotely control the opening and closing of the equipment online;
★ Disinfection effect requirements:
1) Open the equipment for 2 hours in a space of 20m3, and the killing rate of Staphylococcus albus is ≥ 99.90%;
2) Open the equipment for 2 hours in a space of 20m3, and the natural bacteria extinction rate is ≥ 90%;
★ 3) when the active mist ion generation module runs for 15min, the killing rate of the original strain of COVID-19 virus should be greater than 99.90%;
4) When the active mist ion generating module runs for 15min, the killing rate of delta virus strain shall be greater than 99.90%;
★ 5) when the active mist ion generation module runs for 15min, the killing rate of Omicron virus strain shall be greater than 99.90%;
6) When the active mist ion generating module operates for 120min, the killing rate of methicillin-resistant Staphylococcus aureus (MRSA) shall be greater than 95.00% ;
Tags: FCDS-02R White